• <th id="abimu"></th>

      1. 產(chǎn)品中心

        人前列腺素E2(PGE2)酶聯(lián)免疫吸附檢測試劑盒
        ELK9315
        規(guī)格: 價格:
        48T ¥2240.00
        96T ¥3200.00

        Overview 文獻(xiàn)

        Product name: Human PGE2(Prostaglandin E2) ELISA Kit
        Reactivity: Human
        Alternative Names: PG-E2
        Assay Type: Competitive Inhibition
        Sensitivity: 0.97 pg/mL
        Standard: 200 pg/mL
        Detection Range: 3.13-200 pg/mL
        Sample Type: serum, plasma and other biological fluids
        Assay Length: 2h
        Research Area: Metabolic pathway;Infection immunity;Endocrinology;Hormone metabolism;
        Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human PGE2 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human PGE2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human PGE2 in the samples is then determined by comparing the OD of the samples to the standard curve.

        標(biāo)準(zhǔn)曲線

        Concentration (pg/mL) OD Corrected OD
        200.00 0.215
        100.00 0.397
        50.00 0.715
        25.00 0.913
        12.50 1.112
        6.25 1.583
        3.13 1.733
        0.00 2.211

        精密度

        Intra-assay Precision (Precision within an assay):CV%<8%

        Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

        Inter-assay Precision (Precision between assays):CV%<10%

        Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

        回收率

        Matrices listed below were spiked with certain level of recombinant PGE2 and the recovery rates were calculated by comparing the measured value to the expected amount of PGE2 in samples.
        Matrix Recovery range Average
        serum(n=5) 80-96% 88%
        EDTA plasma(n=5) 82-98% 90%
        Heparin plasma(n=5) 80-95% 87%

        線性

        The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PGE2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
        Matrix 1:2 1:4 1:8 1:16
        serum(n=5) 85-94% 87-105% 92-101% 79-90%
        EDTA plasma(n=5) 81-94% 93-107% 93-102% 91-105%
        Heparin plasma(n=5) 92-96% 86-98% 95-101% 87-94%