• <th id="abimu"></th>

      1. 產(chǎn)品中心

        人細(xì)胞角蛋白19片段抗原21-1(CYFRA21-1)酶聯(lián)免疫吸附檢測試劑盒
        ELK1967
        規(guī)格: 價格:
        48T ¥1960.00
        96T ¥2800.00

        Overview

        Product name: Human CYFRA21-1(Cytokeratin Fragment Antigen 21-1) ELISA Kit
        Reactivity: Human
        Alternative Names: CYFRA21-1
        Assay Type: Sandwich
        Sensitivity: 0.117 ng/mL
        Standard: 20 ng/mL
        Detection Range: 0.32-20 ng/mL
        Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
        Assay Length: 3.5h
        Research Area: Tumor immunity;Dermatology;
        Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human CYFRA21-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human CYFRA21-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human CYFRA21-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human CYFRA21-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

        標(biāo)準(zhǔn)曲線

        Concentration (ng/mL) OD Corrected OD
        20.00 2.045 1.951
        10.00 1.502 1.408
        5.00 1.034 0.940
        2.50 0.875 0.781
        1.25 0.539 0.445
        0.63 0.397 0.303
        0.32 0.206 0.112
        0.00 0.094 0.000

        精密度

        Intra-assay Precision (Precision within an assay):CV%<8%

        Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

        Inter-assay Precision (Precision between assays):CV%<10%

        Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

        回收率

        Matrices listed below were spiked with certain level of recombinant CYFRA21-1 and the recovery rates were calculated by comparing the measured value to the expected amount of CYFRA21-1 in samples.
        Matrix Recovery range Average
        serum(n=5) 78-90% 84%
        EDTA plasma(n=5) 84-97% 90%
        Heparin plasma(n=5) 87-95% 91%

        線性

        The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CYFRA21-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
        Matrix 1:2 1:4 1:8 1:16
        serum(n=5) 89-102% 92-101% 88-102% 87-102%
        EDTA plasma(n=5) 85-97% 84-100% 95-103% 88-104%
        Heparin plasma(n=5) 86-96% 91-107% 87-101% 95-102%