SERPINF2; AAP; A2-PI; API; PLI; A2AP; Alpha 2 Antiplasmin; Serpin Peptidase Inhibitor Clade F Member 2; Pigment Epithelium Derived Factor
Assay Type:
Sandwich
Sensitivity:
0.69 ng/mL
Standard:
100 ng/mL
Detection Range:
1.57-100 ng/mL
Sample Type:
Serum, plasma and other biological fluids.
Assay Length:
3.5h
Research Area:
Hematology;
Test principle:
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human a2PI. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human a2PI. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human a2PI, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human a2PI in the samples is then determined by comparing the OD of the samples to the standard curve.
標準曲線
Concentration (ng/mL)
OD
Corrected OD
100.00
2.204
2.121
50.00
1.700
1.617
25.00
0.935
0.852
12.50
0.752
0.669
6.25
0.543
0.460
3.13
0.351
0.268
1.57
0.156
0.073
0.00
0.083
0.000
精密度
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
回收率
Matrices listed below were spiked with certain level of recombinant a2PI and the recovery rates were calculated by comparing the measured value to the expected amount of a2PI in samples.
Matrix
Recovery range
Average
serum(n=5)
83-98%
90%
EDTA plasma(n=5)
80-95%
87%
Heparin plasma(n=5)
86-99%
92%
線性
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of a2PI and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.