HLF2; LF; GIG12; Lactotransferrin; Growth-inhibiting protein 12; Talalactoferrin; Lfcin-H; Kaliocin-1
Assay Type:
Sandwich
Sensitivity:
2.84 ng/mL
Standard:
400 ng/mL
Detection Range:
6.25-400 ng/mL
Sample Type:
Serum, plasma and other biological fluids.
Assay Length:
3.5h
Research Area:
Tumor immunity;Infection immunity;
Test principle:
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat LTF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat LTF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat LTF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat LTF in the samples is then determined by comparing the OD of the samples to the standard curve.
標(biāo)準(zhǔn)曲線
Concentration (ng/mL)
OD
Corrected OD
400.00
2.207
2.118
200.00
1.563
1.474
100.00
1.186
1.097
50.00
0.882
0.793
25.00
0.527
0.438
12.50
0.325
0.236
6.25
0.206
0.117
0.00
0.089
0.000
精密度
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
回收率
Matrices listed below were spiked with certain level of recombinant LTF and the recovery rates were calculated by comparing the measured value to the expected amount of LTF in samples.
Matrix
Recovery range
Average
serum(n=5)
80-97%
88%
EDTA plasma(n=5)
82-95%
88%
Heparin plasma(n=5)
88-99%
93%
線性
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LTF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.