        Overview

        Product name: Human Ab1-40(Amyloid Beta Peptide 1-40) ELISA Kit
        Reactivity: Human
        Alternative Names: Amyloid Beta Peptide 1-40
        Assay Type: Sandwich
        Sensitivity: 4.58 pg/mL
        Standard: 1000 pg/mL
        Detection Range: 15.63-1000 pg/mL
        Sample Type: Serum, plasma and other biological fluids
        Assay Length: 3.5h
        Research Area: Infection immunity;Neuro science;
        Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human Ab1-40. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human Ab1-40. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human Ab1-40, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human Ab1-40 in the samples is then determined by comparing the OD of the samples to the standard curve.

        標(biāo)準(zhǔn)曲線

        Concentration (pg/mL) OD Corrected OD
        1000.00 2.082 1.987
        500.00 1.621 1.526
        250.00 0.997 0.902
        125.00 0.765 0.670
        62.50 0.503 0.408
        31.25 0.351 0.256
        15.63 0.221 0.126
        0.00 0.095 0.000

        精密度

        Intra-assay Precision (Precision within an assay):CV%<8%

        Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

        Inter-assay Precision (Precision between assays):CV%<10%

        Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

        回收率

        Matrices listed below were spiked with certain level of recombinant Ab1-40 and the recovery rates were calculated by comparing the measured value to the expected amount of Ab1-40 in samples.
        Matrix Recovery range Average
        serum(n=5) 80-94% 87%
        EDTA plasma(n=5) 89-101% 95%
        Heparin plasma(n=5) 82-95% 88%

        線性

        The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ab1-40 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
        Matrix 1:2 1:4 1:8 1:16
        serum(n=5) 89-97% 86-102% 94-106% 85-97%
        EDTA plasma(n=5) 85-97% 93-106% 83-95% 88-103%
        Heparin plasma(n=5) 95-103% 88-96% 81-95% 91-102%