        Overview

        Product name: Human ANG1(Angiopoietin1) ELISA Kit
        Reactivity: Human
        Alternative Names: ANG1; AGP1; AGPT; ANG-I; ANGPT1; Angiopoietin 1
        Assay Type: Sandwich
        Sensitivity: 0.11 ng/mL
        Standard: 15 ng/mL
        Detection Range: 0.24-15 ng/mL
        Sample Type: Serum, plasma, tissue homogenates and other biological fluids
        Assay Length: 3.5h
        Research Area: Tumor immunity;Cardiovascular biology;
        Uniprot ID: Q15389
        Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human ANG1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human ANG1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human ANG1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human ANG1 in the samples is then determined by comparing the OD of the samples to the standard curve.

        標(biāo)準(zhǔn)曲線

        Concentration (ng/mL) OD Corrected OD
        15.00 2.035 1.932
        7.50 1.593 1.490
        3.75 1.194 1.091
        1.88 0.831 0.728
        0.94 0.582 0.479
        0.47 0.363 0.260
        0.24 0.215 0.112
        0.00 0.103 0.000

        精密度

        Intra-assay Precision (Precision within an assay):CV%<8%

        Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

        Inter-assay Precision (Precision between assays):CV%<10%

        Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

        回收率

        Matrices listed below were spiked with certain level of recombinant ANG1 and the recovery rates were calculated by comparing the measured value to the expected amount of ANG1 in samples.
        Matrix Recovery range Average
        serum(n=5) 82-94% 88%
        EDTA plasma(n=5) 85-97% 91%
        Heparin plasma(n=5) 81-93% 87%

        線性

        The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ANG1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
        Matrix 1:2 1:4 1:8 1:16
        serum(n=5) 79-92% 80-93% 81-98% 89-102%
        EDTA plasma(n=5) 89-97% 86-95% 81-93% 91-103%
        Heparin plasma(n=5) 85-99% 82-96% 92-101% 82-90%